* Temperature range: -40 °C to +180 °C
* APT.line™ preheating chamber technology
* Intuitive touchscreen controller with time-segment and real-time programming
* Internal data logger, measured values can be read out in open format via USB
* BINDER Multi Management Software APT-COM™ Basic Edition
* Programmable condensation protection for test material
* Troubleshooting system with visual and audible alarms
* CFC-free refrigerant R-452A
* 1 stainless steel rack
* Access port with silicone plug (models 56, 115, 240, 400), 50 mm, left
* 2 access ports with silicone plugs (models 720, 1020), 80 mm, left and right
* Heated viewing window with LED interior lighting
* Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
* 4 castors, two with brakes
* Computer interface: Ethernet
* Adjustable ramp function
* Integrated chart recorder
* Real-time clock
* Door heating
* Inner chamber made of stainless steel
* Cooling with compressor cooling unit
Description MK115-400V * MK115-480V-C *
     
Article Number 9020-0375 9020-0303
Performance Data Temperature    
Temperature range -40…180 -40…180
Temperature variation [± K]    
Temperature variation depending on setpoint [± K] 0.1…2 0.1…2
Temperature fluctuation [± K]    
Temperature fluctuation depending on setpoint [± K] 0.1…0.5 0.1…0.5
Heating-up time from -40 °C to 180 °C [min]    
Average heating-up rate according to IEC 60068-3-5 [K/min] 5.3 5.3
Average heating-up rate according to factory standard [K/min]    
Cooling down time from 180 °C to -40 °C [min] 90 90
Average cooling down time according to IEC 60068-3-5 [K/min] 5 5
Average cooling down time according to factory standard [K/min]    
Max. heat compensation at 20 °C [W]    
Max. heat compensation at 25 °C [W] 1500 1500
Electrical data    
Rated Voltage [V] 400 480
Power frequency [Hz] 50 60
Nominal power [kW] 3.5 3.5
Unit fuse [A] 16 16
Phase (Nominal voltage) 3~ 3~
Measures    
Interior volume [L] 115 115
Net weight of the unit (empty) [kg] 260 195
Permitted load [kg] 60 60
Load per rack [kg] 30 30
Viewing window width [mm] 288 288
Viewing window height [mm] 222 222
Wall clearance back [mm] 300 300
Wall clearance sidewise [mm] 200 200
Internal Dimensions    
Width [mm] 600 600
Height [mm] 480 480
Depth [mm] 400 400
Doors    
Unit doors 1 1
Housing dimensions not incl. fittings and connections    
Width net [mm] 980 980
Height net [mm] 1725 1725
Depth net [mm] 865 865
Environment-specific data    
Energy consumption at 20 °C 650 Wh/h 650 Wh/h
Sound-pressure level [dB(A)] 62 67
Fixtures    
Number of shelves (std./max.) 1/4 1/4