Binder | Model MKF 1020 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control

Download

Binder | Model MKF 1020 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control

Binder

Model MKF 1020 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity ControlBinder Test Cihazları Türkiye Distribütörü | Enfor

* Temperature range: -40 °C to +180 °C
* Humidity range: 10% to 98% RH
* Humidity regulation with capacitive humidity sensor and steam humidification
* APT.line™ preheating chamber technology
* Intuitive touchscreen controller with time-segment and real-time programming
* CFC-free refrigerant R-452A
* Integrated water-storage tank, 20 L
* 1 stainless steel rack
* Access port with silicone plug (models 56, 115, 240, 400), 50 mm, left
* 2 access ports with silicone plugs (models 720, 1020), 80 mm, left and right
* 4 zero-voltage relay contacts
* Heated viewing window with LED interior lighting

 

Description MKF1020-400V * MKF1020-480V-C *
     
Article Number 9020-0409 9020-0448
Performance Data Temperature    
Temperature range -40…180 -40…180
Temperature variation depending on setpoint [± K] 0.1…1.8 0.1…1.8
Temperature fluctuation depending on setpoint [± K] 0.1…0.5 0.1…0.5
Heating-up time from -40 °C to 180 °C [min] 60 60
Average heating-up rate according to IEC 60068-3-5 [K/min] 5.5 5.5
Cooling down time from 180 °C to -40 °C [min] 120 120
Average cooling down time according to IEC 60068-3-5 [K/min] 5 5
Max. heat compensation at 20 °C [W] 6000 6000
Max. heat compensation at 25 °C [W]    
Performance Data Climate    
Temperature range [°C] 10…95 10…95
Temperature variation [± K] 0.1…2 0.1…2
Temperature fluctuation depending on setpoint [± K] 0.1…1.5 0.1…1.5
Humidity range 10…98 10…98
Humidity fluctuation depending on setpoint 2,5 ± % RH 2,5 ± % RH
Dew point temperature range [°C] 5…94 5…94
Max. heat compensation at 25 °C at 90 % RH [W] 1500 1500
Electrical data    
Rated Voltage [V] 400 480
Power frequency [Hz] 50 60
Nominal power [kW] 12.5 12.5
Unit fuse [A] 32 32
Phase (Nominal voltage) 3~ 3~
Measures    
Interior volume [L] 991 991
Net weight of the unit (empty) [kg] 636 636
Permitted load [kg] 200 200
Load per rack [kg] 40 40
Viewing window width [mm] 508 508
Viewing window height [mm] 300 300
Wall clearance back [mm] 300 300
Wall clearance sidewise [mm] 200 200
Internal Dimensions    
Width [mm] 1200 1200
Height [mm] 1020 1020
Depth [mm] 810 810
Doors    
Unit doors 1 1
Housing dimensions not incl. fittings and connections    
Width net [mm] 1580 1580
Height net [mm] 2005 2005
Depth net [mm] 1480 1480
Environment-specific data    
Energy consumption at 25 °C and 60 % RH [Wh/h] 2200 2200
Sound-pressure level [dB(A)] 69 69
Fixtures    
Number of shelves (std./max.) 1/11 1/11