* Temperature range: -40 °C to +180 °C
* Humidity range: 10% to 98% RH
* Humidity regulation with capacitive humidity sensor and steam humidification
* APT.line™ preheating chamber technology
* Intuitive touchscreen controller with time-segment and real-time programming
* CFC-free refrigerant R-452A
* Integrated water-storage tank, 20 L
* 1 stainless steel rack
* Access port with silicone plug (models 56, 115, 240, 400), 50 mm, left
* 2 access ports with silicone plugs (models 720, 1020), 80 mm, left and right
* 4 zero-voltage relay contacts
* Heated viewing window with LED interior lighting

 

Description MKF400-400V * MKF400-480V-C *
     
Article Number 9020-0408 9020-0446
Performance Data Temperature    
Temperature range -40…180 -40…180
Temperature variation depending on setpoint [± K] 0.1…1.2 0.1…1.2
Temperature fluctuation depending on setpoint [± K] 0.1…0.5 0.1…0.5
Heating-up time from -40 °C to 180 °C [min] 55 55
Average heating-up rate according to IEC 60068-3-5 [K/min] 5 5
Cooling down time from 180 °C to -40 °C [min] 115 115
Average cooling down time according to IEC 60068-3-5 [K/min] 5 5
Max. heat compensation at 20 °C [W] 4500 4500
Max. heat compensation at 25 °C [W]    
Performance Data Climate    
Temperature range [°C] 10…95 10…95
Temperature variation [± K] 0.1…1.5 0.1…1.5
Temperature fluctuation depending on setpoint [± K] 0.1…1.3 0.1…1.3
Humidity range 10…98 10…98
Humidity fluctuation depending on setpoint 2,5 ± % RH 2,5 ± % RH
Dew point temperature range [°C] 5…94 5…94
Max. heat compensation at 25 °C at 90 % RH [W] 800 800
Electrical data    
Rated Voltage [V] 400 480
Power frequency [Hz] 50 60
Nominal power [kW] 8 8
Unit fuse [A] 16 16
Phase (Nominal voltage) 3~ 3~
Measures    
Interior volume [L] 417 417
Net weight of the unit (empty) [kg] 432 432
Permitted load [kg] 150 150
Load per rack [kg] 30 30
Viewing window width [mm] 508 508
Viewing window height [mm] 300 300
Wall clearance back [mm] 500 500
Wall clearance sidewise [mm] 300 300
Internal Dimensions    
Width [mm] 735 735
Height [mm] 700 700
Depth [mm] 810 810
Doors    
Unit doors 1 1
Housing dimensions not incl. fittings and connections    
Width net [mm] 1115 1115
Height net [mm] 1710 1710
Depth net [mm] 1400 1400
Environment-specific data    
Energy consumption at 25 °C and 60 % RH [Wh/h] 1900 1900
Sound-pressure level [dB(A)] 65 65
Fixtures    
Number of shelves (std./max.) 1/6 1/6