Binder | Model MKF 56 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control
Binder
Model MKF 56 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control
* Temperature range: -40 °C to +180 °C * Humidity range: 10% to 98% RH * Humidity regulation with capacitive humidity sensor and steam humidification * APT.line™ preheating chamber technology * Intuitive touchscreen controller with time-segment and real-time programming * CFC-free refrigerant R-452A * Integrated water-storage tank, 20 L * 1 stainless steel rack * Access port with silicone plug (models 56, 115, 240, 400), 50 mm, left * 2 access ports with silicone plugs (models 720, 1020), 80 mm, left and right * 4 zero-voltage relay contacts * Heated viewing window with LED interior lighting |
Description | MKF056-230V * | MKF056-240V * |
Article Number | 9020-0378 | 9020-0389 |
Performance Data Temperature | ||
Temperature range | -40…180 | -40…180 |
Temperature variation depending on setpoint [± K] | 0.5…1.5 | 0.5…1.5 |
Temperature fluctuation depending on setpoint [± K] | 0.1…0.5 | 0.1…0.5 |
Heating-up time from -40 °C to 180 °C [min] | ||
Average heating-up rate according to IEC 60068-3-5 [K/min] | 5 | 5 |
Cooling down time from 180 °C to -40 °C [min] | 90 | 90 |
Average cooling down time according to IEC 60068-3-5 [K/min] | 5 | 5 |
Max. heat compensation at 20 °C [W] | ||
Max. heat compensation at 25 °C [W] | 1200 | 1200 |
Performance Data Climate | ||
Temperature range [°C] | 10…95 | 10…95 |
Temperature variation [± K] | ||
Temperature fluctuation depending on setpoint [± K] | 0.1…0.5 | 0.1…0.5 |
Humidity range | 10…98 | 10…98 |
Humidity fluctuation depending on setpoint | ≤2,5 ± % RH | ≤2,5 ± % RH |
Dew point temperature range [°C] | 5…94 | 5…94 |
Max. heat compensation at 25 °C at 90 % RH [W] | 250 | 250 |
Electrical data | ||
Rated Voltage [V] | 200…230 | 200…240 |
Power frequency [Hz] | 50 | 60 |
Nominal power [kW] | 2.8 | 2.8 |
Unit fuse [A] | 16 | 16 |
Phase (Nominal voltage) | 1~ | 1~ |
Measures | ||
Interior volume [L] | 60 | 60 |
Net weight of the unit (empty) [kg] | 175 | 175 |
Permitted load [kg] | 60 | 60 |
Load per rack [kg] | 15 | 15 |
Viewing window width [mm] | 288 | 288 |
Viewing window height [mm] | 255 | 255 |
Wall clearance back [mm] | 300 | 300 |
Wall clearance sidewise [mm] | 200 | 200 |
Internal Dimensions | ||
Width [mm] | 400 | 400 |
Height [mm] | 420 | 420 |
Depth [mm] | 348 | 348 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Width net [mm] | 720 | 720 |
Height net [mm] | 1445 | 1445 |
Depth net [mm] | 780 | 780 |
Environment-specific data | ||
Energy consumption at 25 °C and 60 % RH [Wh/h] | 750 | 750 |
Sound-pressure level [dB(A)] | 59 | 59 |
Fixtures | ||
Number of shelves (std./max.) | 1/4 | 1/4 |