Binder | Model MKFT 240 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control and Extended Low Temperature Range

Download

Binder | Model MKFT 240 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control and Extended Low Temperature Range

Binder

Model MKFT 240 | Dynamic Climate Chamber for Rapid Temperature Changes with Humidity Control and Extended Low Temperature Range

Binder Test Cihazları Türkiye Distribütörü | Enfor

* Temperature range: -70 °C to +180 °C
* Humidity range: 10 % to 98 % RH
* Integrated water-storage tank, 20 liters
* 4 zero-voltage relay contacts
* APT.line™ preheating chamber technology
* Programmable condensation protection for test material
* Intuitive touchscreen controller with time-segment and real-time programming
* Internal data logger, measured values can be read out in open format via USB
* Heated viewing window with LED interior lighting
* Humidity regulation with capacitative humidity sensor and vapor humidification
* BINDER Multi Management Software APT-COM™ Basic Edition
* Troubleshooting system with visual and audible alarms
* Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
* Access port with silicone plug (Models 115, 240: 50 mm, left)
* 2 access ports with silicone plugs (Model 720: 80 mm, left and right)
* 4 castors, two with brakes
* Computer interface: Ethernet
* Adjustable ramp function
* Integrated chart recorder
* Real-time clock
* Door heating
* Alarm notification in the event of insufficient water in fresh water tank
* 1 stainless steel rack
* Complete safety connection kit for water supply and drainage, up to 1 m in height
* Inner chamber made of stainless steel
* CFC-free refrigerants R-452A and R-23
* Cooling with cascade compressor cooling unit
 

 

Description MKFT240-400V * MKFT240-480V-C *
     
Article Number 9020-0383 9020-0361
Performance Data Temperature    
Temperature range -70…180 -70…180
Temperature variation depending on setpoint [± K] 0.2…1.8 0.2…1.8
Temperature fluctuation depending on setpoint [± K] 0.1…0.5 0.1…0.5
Average heating-up rate according to IEC 60068-3-5 [K/min] 5 5
Average cooling down time according to IEC 60068-3-5 [K/min] 4.2 4.2
Max. heat compensation at 25 °C [W] 3000 3000
Performance Data Climate    
Temperature range [°C] 10…95 10…95
Temperature fluctuation depending on setpoint [± K] 0.1…1.5 0.1…1.5
Humidity range 10…98 10…98
Humidity fluctuation depending on setpoint ≤2,5 ± % RH ≤2,5 ± % RH
Dew point temperature range [°C] 5…94 5…94
Max. heat compensation at 25 °C at 90 % RH [W] 400 400
Electrical data    
Rated Voltage [V] 400 480
Power frequency [Hz] 50 60
Nominal power [kW] 6 6
Unit fuse [A] 16 16
Phase (Nominal voltage) 3~ 3~
Measures    
Interior volume [L] 228 228
Net weight of the unit (empty) [kg] 415 579
Permitted load [kg] 70 70
Load per rack [kg] 30 30
Viewing window width [mm] 508 508
Viewing window height [mm] 300 300
Wall clearance back [mm] 300 300
Wall clearance sidewise [mm] 200 200
Internal Dimensions    
Width [mm] 735 735
Height [mm] 700 700
Depth [mm] 443 443
Doors    
Unit doors 1 1
Housing dimensions not incl. fittings and connections    
Width net [mm] 1115 1115
Height net [mm] 1940 1940
Depth net [mm] 925 925
Environment-specific data    
Sound-pressure level [dB(A)] 67 70
Fixtures    
Number of shelves (std./max.) 1/6 1/6