Binder | Model MKT 115 | Dynamic Climate Chamber for Rapid Temperature Changes with Extended Low Temperature Range
Binder
Model MKT 115 | Dynamic Climate Chamber for Rapid Temperature Changes with Extended Low-Temperature Range
* Temperature range: -70 °C to +180 °C * 4 zero-voltage relay contacts * APT.line™ preheating chamber technology * Programmable condensation protection for test material * Heated viewing window with LED interior lighting * BINDER Multi Management Software APT-COM™ Basic Edition * Troubleshooting system with visual and audible alarms * Intuitive touchscreen controller with time-segment and real-time programming * Internal data logger, measured values can be read out in open format via USB * Access port with silicone plug (Models 115, 240: 50 mm, left) * 2 access ports with silicone plugs (Model 720: 80 mm, left and right) * Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm * 4 castors, two with brakes * Computer interface: Ethernet * Adjustable ramp function * Integrated chart recorder * Real-time clock * Door heating * 1 stainless steel rack * Inner chamber made of stainless steel * CFC-free refrigerants R-452A and R-23 * Cooling with cascade compressor cooling unit |
Description | MKT115-400V * | MKT115-480V-C * |
Article Number | 9020-0385 | 9020-0363 |
Performance Data Temperature | ||
Temperature range | -70…180 | -70…180 |
Temperature variation depending on setpoint [± K] | 0.2…1.8 | 0.2…1.8 |
Temperature fluctuation depending on setpoint [± K] | 0.1…0.6 | 0.1…0.6 |
Average heating-up rate according to IEC 60068-3-5 [K/min] | 5.3 | 5.3 |
Cooling down time from 180 °C to -70 °C [min] | 110 | 110 |
Average cooling down time according to IEC 60068-3-5 [K/min] | 4.2 | 4.2 |
Max. heat compensation at 25 °C [W] | 1500 | 1500 |
Electrical data | ||
Rated Voltage [V] | 400 | 480 |
Power frequency [Hz] | 50 | 60 |
Nominal power [kW] | 5.5 | 5.5 |
Unit fuse [A] | 16 | 16 |
Phase (Nominal voltage) | 3~ | 3~ |
Measures | ||
Interior volume [L] | 115 | 115 |
Net weight of the unit (empty) [kg] | 305 | 223 |
Permitted load [kg] | 60 | 60 |
Load per rack [kg] | 30 | 30 |
Viewing window width [mm] | 228 | 228 |
Viewing window height [mm] | 222 | 222 |
Wall clearance back [mm] | 300 | 300 |
Wall clearance sidewise [mm] | 200 | 200 |
Internal Dimensions | ||
Width [mm] | 600 | 600 |
Height [mm] | 480 | 480 |
Depth [mm] | 400 | 400 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Width net [mm] | 980 | 980 |
Height net [mm] | 1725 | 1725 |
Depth net [mm] | 865 | 865 |
Environment-specific data | ||
Sound-pressure level [dB(A)] | 64 | 67 |
Fixtures | ||
Number of shelves (std./max.) | 1/4 | 1/4 |