Binder | Model CB-170 | CO₂ incubator

Download

Binder | Model CB-170 | CO₂ incubator

 

Binder 

Model CB-170 | CO₂ incubator with hot air sterilization and heat sterilizable CO₂ sensor

Binder Test Cihazları Türkiye Distribütörü | Enfor

 

Optimized humidification system with condensation protection
Tried-and-tested anti-contamination concept with 180 °C hot air sterilization
Hot-air sterilizable CO₂ sensor with single-beam infrared technology
Seamless inner chamber made of stainless steel
Fanless interior with flanges as shelf support system
Intuitive touchscreen controller with Venturi CO₂ gas mixing nozzle
Internal data recording can be read out in open format via USB interface
Fault diagnosis system with visual and acoustic alarm
Interfaces: Ethernet, USB, zero-voltage alarm contact
Fail-safe CO₂ system for protecting the pH of cell cultures

 

Description CB170-230V * CB170-230V-D * CB170-230V-O * CB170UL-120V * CB170UL-120V-D * CB170UL-120V-O *
Article Number 9640-0009 9640-0027 9640-0010 9640-0011 9640-0028 9640-0012
Performance Data Temperature            
Temperature range +4 °C above ambient temperature to [°C] 60 60   60 60  
Temperature range +5 °C above ambient temperature to [°C]            
Temperature range +6 °C above ambient temperature to     60 °C     60 °C
Temperature variation at 37 °C [± K] 0.3 0.3 0.4 0.3 0.3 0.4
Temperature fluctuation at 37 °C [± K] 0.1 0.1 0.1 0.1 0.1 0.1
Recovery time after 30 seconds door open at 37 °C [min] 6 6 6 6 6 6
Performance Data Climate            
Humidity range 90…95 90…95 90…95 90…95 90…95 90…95
CO₂-Data            
CO₂ range [Vol.-% CO₂] 0…20 0…20 0…20 0…20 0…20 0…20
CO₂ measurement IR IR IR IR IR IR
CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min] 5 5 5 5 5 5
O₂-Data            
Standard O₂ control ranges: O₂ range [Vol.-% O₂]     0.2…20     0.2…20
O₂ control ranges with option: O₂ range [Vol.-% O₂]     10…95     10…95
O₂ Recovery time after 30 seconds door open at 5 Vol.-% O₂ [min]     10     10
O₂ Recovery time after 30 seconds door open at 10 Vol.-% O₂ [min]            
Electrical data            
Rated Voltage [V] 200…230 200…230 200…230 100…120 100…120 100…120
Power frequency [Hz] 50/60 50/60 50/60 50/60 50/60 50/60
Nominal power [kW] 1.4 1.4 1.4 1.4 1.4 1.4
Unit fuse [A] 10 10 10 16 16 16
Phase (Nominal voltage) 1~ 1~ 1~ 1~ 1~ 1~
Measures            
Interior volume [L] 170 170 170 170 170 170
Net weight of the unit (empty) [kg] 98 98 98 98 98 98
Permitted load [kg] 30 30 30 30 30 30
Load per rack [kg] 10 10 10 10 10 10
Wall clearance back [mm] 100 100 100 100 100 100
Wall clearance sidewise [mm] 50 50 50 50 50 50
Internal Dimensions            
Width [mm] 560 560 560 560 560 560
Height [mm] 600 600 600 600 600 600
Depth [mm] 505 505 505 505 505 505
Doors            
Inner doors 1 1 1 1 1 1
Unit doors 1 1 1 1 1 1
Housing dimensions not incl. fittings and connections            
Width net [mm] 680 680 680 680 680 680
Height net [mm] 870 870 870 870 870 870
Depth net [mm] 715 715 715 715 715 715
Environment-specific data            
Energy consumption at 37 °C [Wh/h] 55 55 55 55 55 55
Sound-pressure level [dB(A)] 41 41 41 41 41 41
Fixtures            
Number of shelves (std./max.) 3/6 3/6 3/6 3/6 3/6 3/6