Binder | Model CB-S 260 | CO₂ incubator

Download

Binder | Model CB-S 260 | CO₂ incubator

Binder 

Model CB-S 260 | CO₂ incubator with hot air sterilization

Binder Test Cihazları Türkiye Distribütörü | Enfor

 

Temperature range: +6 °C above ambient temperature to +50 °C
Humidity range: 90 to 95% RH
CO₂ control range: 0 to 20 vol.% CO₂
Tried-and-tested anti-contamination concept with 180 °C hot air sterilization
Humidification system with condensation protection
CO₂ gas mixing jet with Venturi effect
CO₂ sensor with infrared technology
LCD to display temperature and CO₂ along with additional information and alarms
Internal data logger, measured values can be read out in open format via USB
Tightly fitted inner door made of safety glass
Tightly fitted inner door made of safety glass
Perforated shelves stainless steel
Troubleshooting system with visual and acoustic alarms
Interfaces: Ethernet, USB, zero-voltage alarm contact
Units can be stacked with stacking adapters; the CB-S is suitable for stacking with CB and CBF models of the same size

 

Description CB260-230V1 CB260-230V-O1 CB260UL-120V1 CB260UL-120V-O1
         
Article Number 9640-0013 9640-0014 9640-0015 9640-0016
Performance Data Temperature        
Temperature range +5 °C above ambient temperature to °C 60  °C   60  °C  
Temperature range +6 °C above ambient temperature to   60  °C   60  °C
Temperature variation at 37 °C [± K] 0.4 0.4 0.4 0.4
Temperature fluctuation at 37 °C [± K] 0.1 0.1 0.1 0.1
Recovery time after 30 seconds door open at 37 °C [min] 6 6 6 6
Performance Data Climate        
Humidity range 90…95 90…95 90…95 90…95
CO₂-Data        
CO₂ range [Vol.-% CO₂] 0…20 0…20 0…20 0…20
CO₂ measurement IR IR IR IR
CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min] 5 5 5 5
O₂-Data        
Standard O2 control ranges: O2 range
[Vol.-% O2]
  0.2…20   0.2…20
O2 control ranges with option: O2 range [Vol.-% O2]   10…95   10…95
O2 Recovery time aer 30 seconds door
open at 5 Vol.-% O2 [min]
  15   15
Electrical data        
Rated Voltage [V] 200…230 200…230 100…120 100…120
Power frequency [Hz] 50/60 50/60 50/60 50/60
Nominal power [kW] 1.5 1.5 1.5 1.5
Unit fuse [A] 10 10 16 16
Phase (Nominal voltage) 1~ 1~ 1~ 1~
Measures        
Interior volume [L] 267 267 267 267
Net weight of the unit (empty) [kg] 125 125 125 125
Permitted load [kg] 40 40 40 40
Load per rack [kg] 10 10 10 10
Wall clearance back [mm] 100 100 100 100
Wall clearance sidewise [mm] 50 50 50 50
Internal Dimensions        
Width [mm] 620 620 620 620
Height [mm] 750 750 750 750
Depth [mm] 575 575 575 575
Doors        
Inner doors 1 1 1 1
Unit doors 1 1 1 1
Housing dimensions not incl. fittings and connections        
Width net [mm] 740 740 740 740
Height net [mm] 1020 1020 1020 1020
Depth net [mm] 785 785 785 785
Environment-specific data        
Energy consumption at 37 °C [Wh/h] 65 65 65 65
Sound-pressure level [dB(A)] 41 41 41 41
Fixtures        
Number of shelves (std./max.) 3/8 3/8 3/8 3/8