* Temperature range: +5 °C above ambient temperature to +300 °C * APT.line™ preheating chamber technology * Forced convection * Adjustable exhaust air flap * Controller with timer functions * 2 chrome-plated racks * Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm |
Description | FD023-230V * | FD023UL-120V * |
Article Number | 9010-0194 | 9010-0196 |
Performance Data Temperature | ||
Temperature range +5 °C above ambient temperature to [°C] | 300 | 300 |
Temperature variation at 150 °C [± K] | 2.5 | 2.5 |
Temperature fluctuation at 150 °C [± K] | 0.3 | 0.3 |
Heating-up time to 150 °C [min] | 25 | 25 |
Recovery time after 30 seconds door open at 150 °C [min] | 6 | 6 |
Air change data | ||
Air change (approx.) at 150 °C [x/h] | 64 | 64 |
Electrical data | ||
Rated Voltage [V] | 230 | 120 |
Power frequency [Hz] | 50/60 | 60 |
Nominal power [kW] | 0.8 | 0.8 |
Unit fuse [A] | 10 | 12.5 |
Phase (Nominal voltage) | 1~ | 1~ |
Measures | ||
Interior volume [L] | 20 | 20 |
Net weight of the unit (empty) [kg] | 27 | 27 |
Permitted load [kg] | 25 | 25 |
Load per rack [kg] | 12 | 12 |
Wall clearance back [mm] | 160 | 160 |
Wall clearance sidewise [mm] | 100 | 100 |
Internal Dimensions | ||
Width [mm] | 222 | 222 |
Height [mm] | 330 | 330 |
Depth [mm] | 300 | 300 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Width net [mm] | 435 | 435 |
Height net [mm] | 495 | 495 |
Depth net [mm] | 520 | 520 |
Environment-specific data | ||
Energy consumption at 150 °C [Wh/h] | 300 | 300 |
Fixtures | ||
Number of shelves (std./max.) | 2/4 | 2/4 |