Binder | Model FD 260 | Drying and Heating Chambers with Forced Convection
Binder | Model FD 260 | Drying and Heating Chambers with Forced Convection
* Temperature range: +10 °C above ambient temperature to +300 °C * High temperature accuracy thanks to APT.line™ technology * Forced convection * Controller with LCD display * Electromechanical control of the exhaust air flap * 2 chrome-plated racks * Units up to 115 L are stackable * Class 2 independent temperature safety device (DIN 12880) with visual temperature alarm * Ergonomic handle design * USB port for recording data |
Description | FD260-230V * | FD260UL-240V * |
Article Number | 9010-0309 | 9010-0310 |
Performance Data Temperature | ||
Temperature range +10 °C above ambient temperature to [°C] | 300 | 300 |
Temperature range +12 °C above ambient temperature to [°C] | ||
Temperature variation at 150 °C [± K] | 1.9 | 1.9 |
Temperature fluctuation at 150 °C [± K] | 0.4 | 0.4 |
Heating-up time to 150 °C [min] | 19 | 19 |
Recovery time after 30 seconds door open at 150 °C [min] | 5 | 5 |
Air change data | ||
Air change (approx.) at 100 °C [x/h] | 9 | 9 |
Electrical data | ||
Rated Voltage [V] | 230 | 240 |
Power frequency [Hz] | 50/60 | 60 |
Nominal power [kW] | 2.3 | 2.5 |
Unit fuse [A] | 12.5 | 16 |
Phase (Nominal voltage) | 1~ | 1~ |
Measures | ||
Interior volume [L] | 259 | 259 |
Net weight of the unit (empty) [kg] | 84 | 84 |
Permitted load [kg] | 270 | 270 |
Load per rack [kg] | 40 | 40 |
Wall clearance back [mm] | 160 | 160 |
Wall clearance sidewise [mm] | 100 | 100 |
Internal Dimensions | ||
Width [mm] | 650 | 650 |
Height [mm] | 780 | 780 |
Depth [mm] | 515 | 515 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Width net [mm] | 810 | 810 |
Height net [mm] | 940 | 940 |
Depth net [mm] | 825 | 825 |
Environment-specific data | ||
Energy consumption at 150 °C [Wh/h] | 410 | 410 |
Sound-pressure level [dB(A)] | 43 | 43 |
Fixtures | ||
Number of shelves (std./max.) | 2/8 | 2/8 |