* Temperature range: +10 °C above ambient temperature to +300 °C * Up to 30% lower energy consumption compared to conventional units on the market * High temperature accuracy thanks to APT.line™ technology * Adjustable fan speed * Controller with LCD display and enhanced time functions * Electromechanical control of the exhaust air flap * 2 chrome-plated racks * Units up to 115 liters are stackable * Class 2 integrated independent adjustable temperature safety device (DIN 12880) with visual alarm * Ergonomic handle design * Ethernet interface * USB port for recording data |
Description | FED115-230V * | FED115UL-120V * |
Article Number | 9010-0293 | 9010-0294 |
Performance Data Temperature | ||
Temperature range +10 °C above ambient temperature to [°C] | 300 | 300 |
Temperature range +12 °C above ambient temperature to [°C] | ||
Temperature variation at 150 °C [± K] | 1.2 | 1.2 |
Temperature fluctuation at 150 °C [± K] | 0.3 | 0.3 |
Heating-up time to 150 °C [min] | 18 | 18 |
Recovery time after 30 seconds door open at 150 °C [min] | 4 | 4 |
Air change data | ||
Air change (approx.) at 100 °C [x/h] | 32 | 32 |
Electrical data | ||
Rated Voltage [V] | 230 | 120 |
Power frequency [Hz] | 50/60 | 60 |
Nominal power [kW] | 1.3 | 1.4 |
Unit fuse [A] | 6.3 | 12.5 |
Phase (Nominal voltage) | 1~ | 1~ |
Measures | ||
Interior volume [L] | 116 | 116 |
Net weight of the unit (empty) [kg] | 57 | 57 |
Permitted load [kg] | 150 | 150 |
Load per rack [kg] | 30 | 30 |
Wall clearance back [mm] | 160 | 160 |
Wall clearance sidewise [mm] | 100 | 100 |
Internal Dimensions | ||
Width [mm] | 550 | 550 |
Height [mm] | 550 | 550 |
Depth [mm] | 385 | 385 |
Doors | ||
Unit doors | 1 | 1 |
Housing dimensions not incl. fittings and connections | ||
Width net [mm] | 710 | 710 |
Height net [mm] | 705 | 705 |
Depth net [mm] | 605 | 605 |
Environment-specific data | ||
Energy consumption at 150 °C [Wh/h] | 340 | 340 |
Sound-pressure level [dB(A)] | 43 | 43 |
Fixtures | ||
Number of shelves (std./max.) | 2/5 | 2/5 |