Binder | Model FED 400 | Drying and Heating Chambers with Forced Convection and Enhanced Timer Functions

Download

Binder | Model FED 400 | Drying and Heating Chambers with Forced Convection and Enhanced Timer Functions

Binder | Model FED 400 | Drying and Heating Chambers with Forced Convection and Enhanced Timer Functions

Binder Test Cihazları Türkiye Distribütörü | Enfor

* Temperature range: +5 °C above ambient temperature to +300 °C
* APT.line™ preheating chamber technology
* Adjustable fan speed
* Adjustable exhaust air flap
* Controller with expanded timer functions
* 2 chrome-plated racks
* Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
* Computer interface: RS 422

 

 

Description FED400-400V * FED400UL-208V *
     
Article Number 9010-0216 9010-0217
Performance Data Temperature    
Temperature range +5 °C above ambient temperature to [°C] 300 300
Temperature variation at 150 °C [± K] 3.8 3.8
Temperature fluctuation at 150 °C [± K] 0.7 0.7
Heating-up time to 150 °C [min] 29 29
Recovery time after 30 seconds door open at 150 °C [min] 6 6
Air change data    
Air change (approx.) at 150 °C [x/h] 18 18
Electrical data    
Rated Voltage [V] 400 208
Power frequency [Hz] 50/60 60
Nominal power [kW] 3.4 3.4
Unit fuse [A] 3 x 16 3 x 16
Phase (Nominal voltage) 3~ 3~
Measures    
Interior volume [L] 400 400
Net weight of the unit (empty) [kg] 145 145
Permitted load [kg] 90 90
Load per rack [kg] 35 35
Wall clearance back [mm] 160 160
Wall clearance sidewise [mm] 100 100
Internal Dimensions    
Width [mm] 1000 1000
Height [mm] 800 800
Depth [mm] 510 510
Doors    
Unit doors 2 2
Housing dimensions not incl. fittings and connections    
Width net [mm] 1235 1235
Height net [mm] 1025 1025
Depth net [mm] 765 765
Environment-specific data    
Energy consumption at 150 °C [Wh/h] 1200 1200
Fixtures    
Number of shelves (std./max.) 2/10 2/10