Series KMF | Constant climate chambers with expanded temperature / humidity range
Model
|
Housing dimensions not incl. fittings and connections
W x H x D
[mm] |
Internal dimension
W x H x D
[mm] |
Net weight
[kg] |
Interior volume
[L] |
KMF 115 |
880 x 1050 x 650 |
600 x 483 x 351 |
127 |
102 |
KMF 240 |
930 x 1465 x 800 |
650 x 785 x 485 |
179 |
247 |
KMF 720 |
1250 x 1925 x 890 |
973 x 1250 x 576 |
295 |
700 |
KMF 115 Constant climate chamber with expanded temperature / humidity range BINDER
Technical Data
Average cooling down time according to IEC 60068-3-5 [K/min] 0.5
Average heating-up rate according to IEC 60068-3-5 [K/min] 1.3
Max. heat compensation at 25 °C [W] 150
Temperature range [°C] -10…100
Dew point temperature range [°C] 5…90
Humidity fluctuation depending on setpoint ≤2,5 ± % RH
Humidity range [% RH] 10…98
Temperature fluctuation depending on setpoint [± K] 0.1…0.3
Temperature range [°C] 10…90
Temperature variation depending on setpoint [± K] 0.2…1
Nominal power [kW] 2
Phase (Nominal voltage) 1~
Power frequency [Hz] 50/60
Rated Voltage [V] 200…230
Unit fuse [A] 16
Inner doors 1
Unit doors 1
Depth [mm] 351
Height [mm] 483
Width [mm] 600
Interior volume [L] 102
Load per rack [kg] 30
Net weight of the unit (empty) [kg] 127
permitted load [kg] 100
Depth net [mm] 650
Height net [mm] 1050
Wall clearance back [mm] 100
Wall clearance sidewise [mm] 100
Width net [mm] 880
Energy consumption at 85 °C and 85 % RH [Wh/h] 570
Sound-pressure level [dB(A)] 52
Number of shelves (std./max.) 1/5
KMF 240 Constant climate chamber with expanded temperature / humidity range BINDER
Technical Data
Average cooling down time according to IEC 60068-3-5 [K/min] 0.4
Average heating-up rate according to IEC 60068-3-5 [K/min] 0.8
Max. heat compensation at 25 °C [W] 200
Temperature range [°C] -10…100
Dew point temperature range [°C] 5…90
Humidity fluctuation depending on setpoint ≤2 ± % RH
Humidity range [% RH] 10…98
Temperature fluctuation depending on setpoint [± K] 0.1…0.3
Temperature range [°C] 10…90
Temperature variation depending on setpoint [± K] 0.1…1
Nominal power [kW] 2.1
Phase (Nominal voltage) 1~
Power frequency [Hz] 50/60
Rated Voltage [V] 200…230
Unit fuse [A] 16
Inner doors 1
Unit doors 1
Depth [mm] 485
Height [mm] 785
Width [mm] 650
Interior volume [L] 247
Load per rack [kg] 30
Net weight of the unit (empty) [kg] 179
permitted load [kg] 100
Depth net [mm] 800
Height net [mm] 1465
Wall clearance back [mm] 100
Wall clearance sidewise [mm] 100
Width net [mm] 930
Energy consumption at 85 °C and 85 % RH [Wh/h] 570
Sound-pressure level [dB(A)] 52
Number of shelves (std./max.) 1/9
KMF 720 Constant climate chamber with expanded temperature / humidity range BINDER
Technical Data
Average cooling down time according to IEC 60068-3-5 [K/min] 0.4
Average heating-up rate according to IEC 60068-3-5 [K/min] 0.7
Max. heat compensation at 25 °C [W] 450
Temperature range [°C] -10…100
Dew point temperature range [°C] 5…90
Humidity fluctuation depending on setpoint ≤2 ± % RH
Humidity range [% RH] 10…98
Temperature fluctuation depending on setpoint [± K] 0.1…0.5
Temperature range [°C] 10…90
Temperature variation depending on setpoint [± K] 0.1…1
Nominal power [kW] 3.1
Phase (Nominal voltage) 1~
Power frequency [Hz] 50/60
Rated Voltage [V] 200…230
Unit fuse [A] 16
Inner doors 2
Unit doors 2
Depth [mm] 576
Height [mm] 1250
Width [mm] 973
Interior volume [L] 700
Load per rack [kg] 45
Net weight of the unit (empty) [kg] 295
permitted load [kg] 150
Depth net [mm] 890
Height net [mm] 1925
Wall clearance back [mm] 100
Wall clearance sidewise [mm] 100
Width net [mm] 1250
Energy consumption at 85 °C and 85 % RH [Wh/h] 900
Sound-pressure level [dB(A)] 56
Number of shelves (std./max.) 1/15